PhysicsTheories.com — Radiation Oncology Silo · BED/EQD2 · All Calculators
External-Beam Radiation Dosimetry

Inverse-Square Law & Mayneord F-Factor Calculator

F = [(SSD₂ + dₘ)/(SSD₁ + dₘ)]² · [(SSD₁ + d)/(SSD₂ + d)]²

Convert LINAC output factors and Percent Depth Dose (PDD) tables across Source-to-Surface Distances (SSD) for extended-SSD clinical setups and TBI.

Calculate Extended-SSD Output & Mayneord PDD

Loading calculation...
Why PDD increases with larger SSD

Percent Depth Dose combines two effects as photons travel from dₘₐₓ to depth d: exponential tissue attenuation (TAR/TMR) and geometric divergence (inverse-square falloff). Because depth d - dₘₐₓ is a smaller fraction of SSD₂ + d when SSD₂ > SSD₁, inverse-square drop-off between surface and depth becomes less steep at extended SSD, raising PDD (Mayneord F > 1 for d > dₘₐₓ).

Formula & Variable Table

PDD₂(d, r, SSD₂) ≈ PDD₁(d, r, SSD₁) · F   |   ISL = [(SSD₁ + dₘ)/(SSD₂ + dₘ)]²

SymbolQuantityUnit
SSD₁, SSD₂Reference and target Source-to-Surface Distancescm
dₘₐₓDepth of electronic equilibrium (maximum dose)cm
dCalculation depth below phantom/patient surfacecm
FMayneord F-factor (geometric PDD ratio)dimensionless

Worked Clinical Dosimetry Example

A 6 MV beam (dₘₐₓ = 1.5 cm, PDD = 67.0% at d = 10 cm, 100 cm SSD) is moved to an extended 120 cm SSD while keeping the surface field size fixed.

Step 1 (Inverse-Square Output at dₘₐₓ): ISL = [(100 + 1.5) / (120 + 1.5)]² = (101.5 / 121.5)² = 0.6979 (dose rate at dₘₐₓ falls by 30.2%).

Step 2 (Mayneord F-Factor): F = (121.5 / 101.5)² × (110.0 / 130.0)² = 1.4329 × 0.7160 = 1.0259.

Answer: PDD(10 cm, 120 SSD) = 67.0% × 1.0259 = 68.7%.

Assumptions & Clinical Limits

The Mayneord F-factor assumes Tissue-Maximum Ratio (TMR) is independent of SSD, ignores small changes in phantom scatter when field size is defined at different depths, and becomes approximate at very large SSD extensions or small fields. Note that when collimators remain unchanged during an extended-SSD setup, the geometric field size diverges in proportion to SSD₂/SSD₁, slightly altering the effective field size scatter factor Sₚ.

Related Radiation Oncology Resources

Radiation Oncology Silo HVL & Attenuation Calculator LINAC & Dosimetry Guide Interactive PDD & Bragg Peak

Frequently Asked Questions

Why do higher photon energies produce deeper dₘₐₓ (skin sparing)?

High-energy photons transfer kinetic energy to forward-directed Compton recoil electrons with longer average range in tissue. Absorbed dose builds up beneath the surface until transient charged-particle equilibrium is reached at depth dₘₐₓ.

What is the difference between PDD and TMR?

Percent Depth Dose (PDD) is measured at fixed SSD relative to dₘₐₓ, so it depends on SSD. Tissue-Maximum Ratio (TMR) is measured at fixed source-to-point distance (SAD/isocenter) and is effectively independent of distance for clinical SSDs.